Free software

About OpticalSetup

Copyright © 2026 Luca Genchi and contributors.

OpticalSetup is free software: you can redistribute it and/or modify it under the terms of the GNU General Public License as published by the Free Software Foundation, either version 3 of the License, or (at your option) any later version.

It is distributed in the hope that it will be useful, but without any warranty; without even the implied warranty of merchantability or fitness for a particular purpose. See the GNU General Public License for more details.

The numbers OpticalSetup reports come from models with stated limits; read each component's page in the wiki before relying on one.

If you use OpticalSetup in your research, teaching or publications, please cite the software:

Bertoncini, A., & Genchi, L. (2026). OpticalSetup - the optical workbench software [Computer software]. Zenodo. https://doi.org/10.5281/zenodo.23191537

Calculators / Fluorescence

Fluorescence

Choose linear (one-photon excited) or nonlinear (two-photon excited) fluorescence, a source (an LED, a CW laser or a pulsed laser; pulsed only for two-photon) and a fluorophore, then emission filters and a detector. Read how strongly the source excites it, how much of its emission is detected, how well the source light is blocked, the crosstalk from a second label, and the excitation rate and saturation of one molecule.

1Spectra, filters and the detected emission
Excitation and emission are scaled to their peaks; the filter stack and the detector QE are fractions. The dark area is the emission that is detected; the dashed vertical line is the source’s central wavelength, and an LED’s or a pulse’s spectrum is drawn as its own curve.
Data table
2Blocking of the filter stack
Optical density −log₁₀T of the whole stack, shown up to 12. The source’s wavelengths need a high value: scattered excitation light can be far stronger than the fluorescence.
Data table

Filters multiply: the stack passes the product of their transmissions.

Fluorescent marker additional information

Excitation rate and saturation of one molecule

From the laser power at the sample: the intensity in a focus or a widefield field, how often one molecule absorbs, how close it is to saturation, and how many of its photons reach the detector [3][5].

    Model
    Measured one-photon and two-photon spectra; one- and two-photon excitation by short pulses per pulse; Gaussian LED bands and transform-limited pulses; generic filters with logistic edges, or your measured curves; typical detector QE curves; a two-level model for saturation, steady under CW and LED light, per pulse under short pulses.
    Evidence
    Equations checked against an independent Python reference (Airy-pattern quadrature, an integrated rate equation, exact filter integrals) and against the detectors’ datasheet values: validation table. The spectra are FPbase’s; they are not checked here.
    Data
    71 fluorescent proteins from FPbase [1][2], each credited to its original paper. Dyes follow once their redistribution terms are confirmed.
    Code
    fluorescence.js, a workbench module the canvas’s specimens can use, and the calculator module.

    Excitation by the source

    A fluorophore absorbs a photon of wavelength λ with a probability set by its decadic molar extinction coefficient ε(λ), the number spectroscopists quote. For one molecule it is an absorption cross-section σ [3]:

    σ(λ)=ln⁡10⋅103 cm3 L−1NA ε(λ)≈3.82×10−21 cm2×ε(λ)M−1 cm−1\sigma(\lambda)=\frac{\ln 10\cdot 10^{3}\ \mathrm{cm^3\,L^{-1}}}{N_A}\,\varepsilon(\lambda)\approx 3.82\times10^{-21}\ \mathrm{cm^2}\times\frac{\varepsilon(\lambda)}{\mathrm{M^{-1}\,cm^{-1}}}

    FPbase lists ε at the absorption peak λpeak, and an excitation spectrum X(λ): the fluorescence measured while the excitation wavelength is scanned. For one emitting species the excitation and absorption spectra have the same shape, because the quantum yield does not depend on which absorption band was excited (Kasha–Vavilov) [3]. The value at a CW laser line λ0 therefore scales ε:

    ε(λ0)=εpeak X(λ0)X(λpeak),B=εpeak ΦF,B(λ0)=ε(λ0) ΦF\varepsilon(\lambda_0)=\varepsilon_{\mathrm{peak}}\,\frac{X(\lambda_0)}{X(\lambda_{\mathrm{peak}})},\qquad B=\varepsilon_{\mathrm{peak}}\,\Phi_F,\qquad B(\lambda_0)=\varepsilon(\lambda_0)\,\Phi_F

    An LED and a pulsed laser have a spectrum w(λ), power per nm. The page takes the same shapes the canvas’s LED source and pulsed laser emit: an LED is a Gaussian band of FWHM Δλ, and a pulse is transform-limited and Gaussian, so its bandwidth follows from its duration τp [14]. Since photons per nm go as w(λ)·λ, the relative excitation, and with it ε and σ, is then averaged over the source’s photons:

    Δλpulse=0.441 λ02c τp,Xˉ=∫w(λ) λ X(λ) dλX(λpeak)∫w(λ) λ dλ\Delta\lambda_{\mathrm{pulse}}=\frac{0.441\,\lambda_0^2}{c\,\tau_p},\qquad \bar X=\frac{\int w(\lambda)\,\lambda\,X(\lambda)\,d\lambda}{X(\lambda_{\mathrm{peak}})\int w(\lambda)\,\lambda\,d\lambda}

    X̄, or X(λ0)/X(λpeak) for a line, is the “relative excitation” the page reports, and ΦF is the fluorescence quantum yield. The molecular brightness B compares fluorophores at their own peaks; B at the source compares them under the light you have. The excitation curve in the first graph is X(λ) itself, so the best of the sources you have can be read off it. A few proteins have only an absorption spectrum on FPbase; it is used the same way.

    Two-photon excitation

    Choosing nonlinear fluorescence switches to two-photon excitation: the molecule absorbs two near-infrared photons at once, with a probability that goes as the square of the intensity, and then emits as after one-photon excitation, so its emission spectrum and quantum yield are unchanged [15][16]. Its strength is the two-photon cross-section σ₂, in GM (1 GM = 10⁻⁵⁰ cm⁴ s per photon), and its spectrum is measured against the laser wavelength; the first graph draws it there, on the same axis as the emission. It is not the one-photon spectrum at twice the wavelength: EGFP’s two-photon peak is near 927 nm, not 976 nm [13]. Only pulsed lasers are offered, and the list holds the 36 proteins for which FPbase has a two-photon spectrum, 20 of them with the peak cross-section that absolute rates need. FPbase records the source of six of these curves (Molina et al. [17]); for the rest it names none, though for fourteen proteins its references include Drobizhev et al. [13].

    The relative excitation is the two-photon curve averaged over the pulse’s photons, and σ₂ at the laser scales from its peak value. Strictly, a broad pulse’s two-photon absorption depends on its second-harmonic spectrum; the average is a fair estimate while the pulse is narrow next to the two-photon spectrum, and pulses wider than 5 % of the wavelength (about 20 fs at 920 nm) are declined. For Gaussian pulses of FWHM τp and photon fluence F, a ground-state molecule at the centre of the focus is driven to n excitations per pulse, and the pulse train then responds as in the one-photon case below:

    n=σ2 gp2 F2τp,gp=2ln⁡2π≈0.664,p=1−e−nn=\frac{\sigma_2\,g_p}{2}\,\frac{F^2}{\tau_p},\qquad g_p=\sqrt{\frac{2\ln 2}{\pi}}\approx0.664,\qquad p=1-e^{-n}

    Here gp is the temporal factor of a Gaussian pulse, ∫I(t)²dt = gp(∫I dt)²/τp [15]. σ₂ is the two-photon absorption cross-section, defined by the photons a thin sample removes from the beam; since each excitation takes two of them, the excitation rate is ½σ₂φ². That is the convention of the reference standards against which the FPbase values were measured [18][19]. At the same average power the signal grows as 1/(frepτp), which is why the repetition rate and the pulse duration both matter; n = 1 marks saturation. Crosstalk compares σ₂·Φ·η of the two labels, which both scale with the square of the intensity.

    Filters, detector and the detected fraction

    The emission spectrum E(λ) passes a stack of filters, whose transmissions multiply, and reaches a detector of quantum efficiency Q(λ). The page reports the fraction of the emitted photons the filters pass, and the fraction that is then counted, over the whole emission spectrum:

    T(λ)=∏iTi(λ),ηT=∫E(λ) T(λ) dλ∫E(λ) dλT(\lambda)=\prod_i T_i(\lambda),\qquad \eta_T=\frac{\int E(\lambda)\,T(\lambda)\,d\lambda}{\int E(\lambda)\,d\lambda}
    ηD=∫E(λ) T(λ) Q(λ) dλ∫E(λ) dλ,Qˉ=ηDηT\eta_D=\frac{\int E(\lambda)\,T(\lambda)\,Q(\lambda)\,d\lambda}{\int E(\lambda)\,d\lambda},\qquad \bar Q=\frac{\eta_D}{\eta_T}

    E(λ) is taken per unit wavelength, as FPbase stores it. A generic filter has peak transmission Tp in its passband and 10−OD outside it, joined by logistic edges whose 10–90 % width Δ you set (0 is an ideal step). Bandpass and notch filters put their edges at the centre ± half the width, so a “525/50” bandpass is centre 525 nm, width 50 nm. The width must be at least three edge widths: narrower, the two edges overlap and the band reaches neither its peak transmission nor its FWHM; from three edge widths on, both hold to 0.3 %:

    Ti(λ)=Tb+(Tp−Tb) S(λ),Tb=10−OD,s(λ;λe)=[1+e−2ln⁡9 (λ−λe)/Δ]−1T_i(\lambda)=T_b+(T_p-T_b)\,S(\lambda),\quad T_b=10^{-\mathrm{OD}},\qquad s(\lambda;\lambda_e)=\Big[1+e^{-2\ln 9\,(\lambda-\lambda_e)/\Delta}\Big]^{-1}
    SLP=s(λ;λe),SSP=1−s(λ;λe)S_{\mathrm{LP}}=s(\lambda;\lambda_e),\qquad S_{\mathrm{SP}}=1-s(\lambda;\lambda_e)
    SBP=s(λ;c−W2)[1−s(λ;c+W2)],Snotch=1−SBPS_{\mathrm{BP}}=s\big(\lambda;c-\tfrac W2\big)\Big[1-s\big(\lambda;c+\tfrac W2\big)\Big],\qquad S_{\mathrm{notch}}=1-S_{\mathrm{BP}}

    The emission side of a dichroic beamsplitter acts like a longpass with weak blocking (OD 1–2). A filter you have measured, or downloaded from its maker, can be loaded as a two-column file (wavelength in nm, then transmission); it is read in your browser and never uploaded, and outside the file’s range its transmission is taken as zero. Its values are read as the file’s header states (a “%” or “percent” in it), otherwise as percent when one exceeds 1.5; choose fractions or percent to override that. The laser line itself is reported as the stack’s transmission and optical density OD = −log10T at the laser wavelength. Back-scattered and reflected laser light can be many orders of magnitude stronger than the fluorescence, which is why emission filters block the line at OD 6 or more.

    The detector curves are typical ones, one representative product each, drawn by OpticalSetup from the manufacturer’s datasheet. Where the datasheet tabulates the photocathode’s radiant sensitivity S, the quantum efficiency follows exactly; the points in between are read off its typical-response figure [9]:

    Q(λ)=S(λ) hce λ=S [A/W]×1239.84λ [nm]Q(\lambda)=\frac{S(\lambda)\,hc}{e\,\lambda}=\frac{S\,[\mathrm{A/W}]\times 1239.84}{\lambda\,[\mathrm{nm}]}
    DetectorRepresentative productDatasheet values reproducedSource
    Bialkali PMT (super bialkali, borosilicate window)Hamamatsu H10720/H10721-110 photosensor module110 mA/W at its 400 nm peak, 34 % QE; 230–700 nm[6]
    GaAsP PMTHamamatsu H7422-40 photosensor module108 mA/W at 420 nm, 176 mA/W at 550 nm, 40 % QE at the 580 nm peak; 300–720 nm[7]
    GaAs PMTHamamatsu H7422-50 photosensor module15 mA/W at 420 nm, 50 mA/W at 550 nm, 90 mA/W at the 800 nm peak; 380–890 nm[7]
    Back-illuminated sCMOS cameraHamamatsu ORCA-Fusion BT (C15440-20UP)95 % QE at 550 nm; 300–1100 nm[8]

    A real detector of the same kind can differ by several percentage points, more near the ends of its range, and with its window, temperature and age. For your own detector, load its measured curve, or enter a constant QE.

    A second fluorophore

    With a second label in the sample, the page evaluates it with the same laser and detection path: how strongly the laser excites it, how much of its emission is detected, and its signal relative to the main fluorophore at equal concentrations,

    X=ε2(λL) ΦF,2 ηD,2ε1(λL) ΦF,1 ηD,1X=\frac{\varepsilon_2(\lambda_L)\,\Phi_{F,2}\,\eta_{D,2}}{\varepsilon_1(\lambda_L)\,\Phi_{F,1}\,\eta_{D,1}}

    X is the bleed-through of label 2 into this channel. The collection geometry is the same for both and cancels. It ignores energy transfer between the labels and any difference in their environment.

    Excitation rate and saturation

    For a diffraction-limited focus the page takes the peak of the Airy pattern of a uniformly filled pupil, I(v) = I0[2J1(v)/v]² with v = 2π·NA·r/λ, whose total power fixes I0 [4]; widefield illumination is a uniform field of diameter D. An LED cannot be focused to a diffraction-limited spot, its light being spatially incoherent, so with an LED only widefield is offered. For pulses, P and I are averages over time. A short pulse has a broad spectrum, and each wavelength focuses to its own Airy peak, π NA² P(λ)/λ²; at the centre of the focus the page therefore weights the cross-section by that, which changes it from the average at λ0 by the factor κ below. κ is 1 for a line; for a broad band it depends on where the absorption sits within it: 1.278 for EGFP excited by a 5 fs pulse at 560 nm, whose absorption lies on the band’s blue side, but 0.893 for mCherry excited by a 5 fs pulse at 488 nm, whose absorption lies on the red side:

    I0=π NA2 Pλ2  (focus),I=4PπD2  (widefield)I_0=\frac{\pi\,\mathrm{NA}^2\,P}{\lambda^2}\ \ \text{(focus)},\qquad I=\frac{4P}{\pi D^2}\ \ \text{(widefield)}
    κ=λ02 ∫w(λ) X(λ) λ−1 dλ∫w(λ) X(λ) λ dλ\kappa=\lambda_0^2\,\frac{\int w(\lambda)\,X(\lambda)\,\lambda^{-1}\,d\lambda}{\int w(\lambda)\,X(\lambda)\,\lambda\,d\lambda}

    One molecule is treated as a two-level system with fluorescence lifetime τ [3][5]. Under a CW laser or an LED it absorbs at the rate k = σφ, where φ is the photon flux (with the source’s mean photon energy for a band). In the steady state a fraction N1 of the time is spent excited, and the emission rate R saturates once kτ approaches 1:

    φ=Iλhc,k=σ φ,N1=kτ1+kτ\varphi=\frac{I\lambda}{hc},\qquad k=\sigma\,\varphi,\qquad N_1=\frac{k\tau}{1+k\tau}
    R=ΦF k1+kτ,Isat=hcλ σ τR=\Phi_F\,\frac{k}{1+k\tau},\qquad I_{\mathrm{sat}}=\frac{hc}{\lambda\,\sigma\,\tau}

    Pulses are taken much shorter than the lifetime and than the period between them (τp at most τ/10 and 1/(10 frep), otherwise the page declines). During a pulse the molecule then does not decay, so a pulse of photon fluence F excites it with probability p = 1 − e−σF; until the next pulse, a period T = 1/frep later, it decays by e−T/τ. In the periodic steady state the fraction excited after each pulse N+ and the emission rate follow, and σF = 1 marks saturation. The signal depends on the energy per pulse, not only on the average power, which is why the repetition rate is asked for: at the same average power, pulsed excitation saturates much sooner than CW. Far below saturation both give the same rate:

    F=Ifrep hν,p=1−e−σF,N+=p1−(1−p) e−T/τF=\frac{I}{f_{\mathrm{rep}}\,h\nu},\qquad p=1-e^{-\sigma F},\qquad N_+=\frac{p}{1-(1-p)\,e^{-T/\tau}}
    R=frep ΦF N+(1−e−T/τ),Ppeak=2ln⁡2/π Pfrep τpR=f_{\mathrm{rep}}\,\Phi_F\,N_+\big(1-e^{-T/\tau}\big),\qquad P_{\mathrm{peak}}=2\sqrt{\ln 2/\pi}\ \frac{P}{f_{\mathrm{rep}}\,\tau_p}

    Of the photons emitted, an isotropic emitter sends the fraction ηcoll into an objective of numerical aperture NA in a medium of index n; the detected photon rate per molecule multiplies the emission rate by that and by the spectral fraction ηD from above:

    ηcoll=1−cos⁡θ2,sin⁡θ=NAn,Rdet=R ηcoll ηD\eta_{\mathrm{coll}}=\frac{1-\cos\theta}{2},\quad \sin\theta=\frac{\mathrm{NA}}{n},\qquad R_{\mathrm{det}}=R\,\eta_{\mathrm{coll}}\,\eta_D

    The page also gives the average power at the sample that would bring kτ, or σF, to 1 in the chosen geometry. In practice the signal of a real fluorophore stops growing in proportion to the power well before that, because triplet and other dark states fill up and the molecule bleaches; the two-level numbers are an upper bound on the rate, not a prediction of the image.

    Every parameter

    SymbolQuantityUnitMeaning
    —Fluorescence process—Linear (one-photon excited) or nonlinear (two-photon excited); two-photon takes a pulsed laser and the markers with a two-photon spectrum.
    λ0Central wavelengthnmVacuum wavelength of the source: 250–1000 nm for one-photon, 550–1400 nm for two-photon excitation. A CW laser’s linewidth is neglected.
    σ2Two-photon cross-sectionGMAt the peak FPbase lists, scaled by the two-photon curve to the laser; 1 GM = 10⁻⁵⁰ cm⁴ s.
    ΔλLED bandwidthnmFWHM of the LED’s Gaussian band, 5–200 nm, the canvas LED’s range.
    τpPulse durationfsIntensity FWHM of transform-limited Gaussian pulses, 5 fs–1 ns.
    frepRepetition rateMHzPulses per second (rate panel, pulsed sources only).
    X(λ)Excitation spectrum—FPbase’s curve, 1 nm samples, linear between them; zero outside the measured range.
    εExtinction coefficientM⁻¹ cm⁻¹At the absorption peak, as FPbase lists it; scaled by X to the laser.
    ΦFQuantum yield—Photons emitted per photon absorbed, from FPbase.
    τFluorescence lifetimensFrom FPbase when listed, or entered; only the saturation panel uses it.
    E(λ)Emission spectrum—FPbase’s curve, taken per unit wavelength.
    c, WCentral wavelength and width of a bandnmBandpass passband or notch stop band, between its half-maximum points; W ≥ 3Δ.
    λeEdge wavelengthnmWhere a longpass or shortpass transmits half its peak.
    ΔEdge widthnm10 %–90 % width of each edge, 0–50 nm; 0 is an ideal step.
    TpPeak transmission%Transmission inside the passband, 1–100 %.
    ODBlocking—Outside the passband T = 10−OD, OD 0–10, below Tp.
    Q(λ)Detector quantum efficiency%Typical curve, constant value or your measured curve; 100 % with no detector (the default).
    PAverage power at the samplemWAverage power after the objective.
    NA, nNumerical aperture, sample index—Focus size and collection cone; NA must not exceed n.
    DField diameterµmDiameter of a uniform widefield field.

    The fluorophore library

    The spectra and numbers come from FPbase, the community-edited fluorescent protein database [1]. FPbase states that its data are free of copyright restrictions, for commercial and non-commercial use, with credit to the original authors [2]; each protein’s file names the paper FPbase credits for the protein and, where FPbase records one, the paper its curves come from, listed below. An entry with neither is left out. They are measurements, not modelled. Each curve is kept at FPbase’s 1 nm sampling: negative values (baseline noise) are set to zero and the curve is scaled to its peak. The excitation peak is taken within 5 nm of the absorption maximum FPbase lists, the wavelength its ε refers to, so a protein’s 280 nm aromatic band can rise above it.

    A record is left out when a listed peak lies more than 5 nm from its own curve’s, because the numbers and the curve then disagree: DsRed and mRuby2 (emission 11 and 7 nm apart) are left out for that reason, and GCaMP6f because its record lists peaks of 300 and 800 nm. Proteins for which FPbase stores another protein’s curve are left out too, since the page would show a relative’s measurement: mEGFP and mEmerald (EGFP’s curves), mCitrine and YPet (Citrine’s), EYFP and Venus (Citrine’s emission), Cerulean (ECFP’s) and mClover3 (Clover’s). Calcium indicators appear in both of their states. Dye spectra (Alexa Fluor, ATTO, cyanine and other dyes) are left out for now: on FPbase most record no source, or only a manufacturer, so their redistribution terms are being confirmed first.

    Other collections were checked and not used: PhotochemCAD [10] and the MPI Brain Research spectra viewer [11] state no licence for their spectra, and PubSpectra [12] has none recorded and dates from 2006. Manufacturers’ spectra viewers do not grant redistribution.

    All 71 fluorophores
    FluorophorePeaks ex / em (nm)ε (M⁻¹ cm⁻¹)ΦFτ (ns)Two-photon peak (nm, σ₂)Original data
    Blue and UV-excited (5)
    Sirius355 / 42415,0000.24——Tomosugi et al. (2009)
    EBFP2383 / 44832,0000.563751, 13.7 GMAi et al. (2007)
    mTagBFP2399 / 45450,6000.642.6—Subach et al. (2011)
    TagBFP402 / 45752,0000.632.6807, 12.1 GMSubach et al. (2008)
    T-Sapphire399 / 51144,0000.62.93—Zapata-Hommer & Griesbeck (2003)
    Cyan (4)
    mTurquoise2434 / 47430,0000.934870Goedhart et al. (2012)
    mCerulean3433 / 47540,0000.874.1860Markwardt et al. (2011)
    ECFP434 / 47732,5000.43866, 23.2 GMHeim & Tsien (1996)
    mTFP1462 / 49264,0000.853.2868, 74 GMAi et al. (2006)
    Green (12)
    StayGold496 / 505159,0000.93——Hirano et al. (2022)
    EGFP488 / 50755,9000.62.6927, 39.6 GMCormack et al. (1996)
    mBaoJin500 / 508128,0000.933.1—Zhang et al. (2024)
    avGFP395 / 50925,0000.79—800Prasher et al. (1992)
    mWasabi493 / 50970,0000.8—926, 42.9 GMAi et al. (2008)
    AausFP1504 / 510170,0000.97——Lambert et al. (2019)
    mStayGold499 / 510164,0000.832.79—Ando et al. (2023)
    Superfolder GFP485 / 51083,3000.653.1932Pédelacq et al. (2005)
    mGreenLantern503 / 514101,8000.72——Campbell et al. (2020)
    Clover505 / 515111,0000.763.2990Lam et al. (2012)
    mNeonGreen506 / 517116,0000.83.1680Shaner et al. (2013)
    Gamillus (on state)504 / 51983,0000.9——Shinoda et al. (2018)
    Yellow (2)
    mVenus515 / 527104,0000.642.7—Kremers et al. (2006)
    Citrine516 / 52977,0000.763.6970, 9.66 GMGriesbeck et al. (2001)
    Orange (3)
    mKOκ551 / 563105,0000.61—1044, 41.7 GMTsutsui et al. (2008)
    mKO2551 / 56563,8000.62—770Sakaue-Sawano et al. (2008)
    mOrange2549 / 56558,0000.62.71080Shaner et al. (2008)
    Red (13)
    tdTomato554 / 581138,0000.69—1052, 140 GMShaner et al. (2004)
    TagRFP555 / 584100,0000.482.31051, 93.8 GMMerzlyak et al. (2007)
    TagRFP-T555 / 58481,0000.412.3—Shaner et al. (2008)
    DsRed-Express2554 / 59135,6000.42——Strack et al. (2008)
    mApple568 / 59275,0000.492.9—Shaner et al. (2008)
    mRuby3558 / 592128,0000.45——Bajar et al. (2016)
    mScarlet3569 / 592104,0000.754—Gadella et al. (2023)
    mScarlet-I569 / 593104,0000.543.1720Bindels et al. (2016)
    mScarlet569 / 594100,0000.73.9740Bindels et al. (2016)
    mStrawberry574 / 59690,0000.29—702Shaner et al. (2004)
    FusionRed580 / 60894,5000.191.8—Shemiakina et al. (2012)
    mCherry587 / 61072,0000.221.41086, 27.6 GMShaner et al. (2004)
    mCherry2589 / 61079,4000.22——Shen et al. (2017)
    Far-red (5)
    mKate2588 / 63362,5000.42.51114, 74.5 GMShcherbo et al. (2009)
    mNeptune2.5599 / 64395,0000.24——Chu et al. (2014)
    mPlum590 / 64941,0000.1—1117, 24.2 GMWang et al. (2004)
    mMaroon1609 / 65780,0000.11—1100Bajar et al. (2016)
    mCardinal604 / 65987,0000.19—1110Chu et al. (2014)
    Near-infrared (8)
    emiRFP670642 / 67087,4000.141.53—Matlashov et al. (2020)
    iRFP670643 / 670114,0000.11—760Shcherbakova & Verkhusha (2013)
    miRFP670642 / 67087,4000.141.42—Shcherbakova et al. (2016)
    smURFP642 / 670180,0000.181.4—Rodriguez et al. (2016)
    miRFP680661 / 68094,0000.1451.67—Matlashov et al. (2020)
    iRFP713690 / 713105,0000.06——Filonov et al. (2011)
    miRFP713690 / 71399,0000.070.89—Matlashov et al. (2020)
    miRFP720702 / 72098,0000.061——Shcherbakova et al. (2018)
    Large Stokes shift (3)
    mAmetrine406 / 52645,0000.58—796, 59.5 GMAi et al. (2008)
    LSSmOrange437 / 57252,0000.45—880Shcherbakova et al. (2012)
    mBeRFP446 / 61165,0000.272890Yang et al. (2013)
    Photoactivatable, photoconvertible and photoswitchable (10)
    rsEGFP2 (on state)478 / 50361,3000.3——Grotjohann et al. (2012)
    mMaple3 (green form)491 / 50615,7600.37——Wang et al. (2014)
    Dendra2 (green form)490 / 50745,0000.53.3—Gurskaya et al. (2006)
    mEos3.2 (green form)507 / 51663,4000.84——Zhang et al. (2012)
    PA-GFP (activated)504 / 51717,4000.79——Patterson (2002)
    Dronpa (on state)503 / 51895,0000.853.6—Ando (2004)
    Dendra2 (red form)553 / 57335,0000.554.4—Gurskaya et al. (2006)
    mEos3.2 (red form)572 / 58032,2000.55——Zhang et al. (2012)
    mMaple3 (red form)568 / 58323,9700.52——Wang et al. (2014)
    PAmCherry1 (activated)564 / 59518,0000.46——Subach et al. (2009)
    Calcium indicators (6)
    jRGECO1a (Ca²⁺-bound)562 / 58863,5540.22—1056, 31 GMDana et al. (2016); spectra: Molina et al. (2019)
    R-GECO1 (Ca²⁺-bound)563 / 58856,4160.2—1055, 31 GMspectra: Molina et al. (2019)
    jRCaMP1a (Ca²⁺-bound)572 / 59461,9200.54—1076, 23 GMspectra: Molina et al. (2019)
    jRCaMP1a (Ca²⁺-free)573 / 59542,7350.29—1076, 25 GMspectra: Molina et al. (2019)
    jRGECO1a (Ca²⁺-free)577 / 5954,7410.12—1076, 18 GMDana et al. (2016); spectra: Molina et al. (2019)
    R-GECO1 (Ca²⁺-free)576 / 5995,1380.15—1072, 24 GMspectra: Molina et al. (2019)

    Precision and limits

    The arithmetic is checked against an independent Python reference to 10⁻⁹ for algebra and exact integrals, 10⁻⁵ for the focus, 2×10⁻⁵ for logistic filter edges and 10⁻⁴ for averages over a source band and for pulse trains, where the page integrates by the trapezoidal rule on 0.05 nm (finer for narrow bands) and splits at hard edges so that ideal steps are exact. The pulse-train check integrates the rate equation through finite pulses; the per-pulse model differs from it by about the ratio of pulse duration to lifetime. The detector curves reproduce their datasheets’ tabulated values to the third decimal. These checks say the page evaluates its model correctly; the inputs set its accuracy. Published ε and ΦF of one protein often differ by 10–20 % between papers; spectra depend on pH, temperature and environment, and FPbase does not record whether an emission curve was measured in photons or in energy per wavelength, which matters a little for broad emission. Read the fractions as good to a few percent at best.

    Not modelled:

    • Three-photon excitation, and two-photon excitation of dyes, for which no openly licensed two-photon spectra were found.
    • Two-photon absorption of a broad pulse beyond the average over its spectrum, and one-photon absorption at a two-photon laser’s wavelength.
    • Pulses not much shorter than the lifetime, chirped pulses, stimulated emission, triplet and dark states, photobleaching and photoswitching kinetics.
    • An excitation filter in front of an LED, and an LED’s real, often asymmetric, spectrum: the page uses a Gaussian band.
    • Concentration effects: inner-filter absorption, reabsorption of the emission, quenching and energy transfer between labels.
    • Polarization, dipole orientation, emission near interfaces, the objective’s own transmission and the dichroic’s reflection of the laser.
    • Filter properties beyond the generic shape: angle of incidence, ripple, out-of-band leaks. Load a measured curve for those.
    • Vectorial high-NA focusing, which makes the focus somewhat wider and its peak lower than the scalar Airy value.

    References

    1. T. J. Lambert, “FPbase: a community-editable fluorescent protein database,” Nat. Methods 16, 277–278 (2019)
    2. FPbase, License, Policies, and Disclaimers (data free of copyright restrictions; credit the original authors)
    3. J. R. Lakowicz, Principles of Fluorescence Spectroscopy, 3rd ed., Springer (2006), chapters 1–2
    4. M. Born and E. Wolf, Principles of Optics, 7th ed., Cambridge University Press (1999), section 8.5.2: the Airy pattern
    5. J. B. Pawley (ed.), Handbook of Biological Confocal Microscopy, 3rd ed., Springer (2006): chapter 16, R. Y. Tsien, L. Ernst and A. Waggoner, “Fluorophores for confocal microscopy: photophysics and photochemistry”
    6. Hamamatsu Photonics, Photomultiplier tube modules H10720/H10721 series, datasheet TPMO1062E
    7. Hamamatsu Photonics, Photosensor modules H7422 series, datasheet (copy hosted by UC San Diego)
    8. Hamamatsu Photonics, ORCA-Fusion BT digital CMOS camera C15440-20UP, catalogue SCAS0143J
    9. Hamamatsu Photonics, Photomultiplier Tubes: Basics and Applications, 4th ed. (2017): radiant sensitivity and quantum efficiency
    10. M. Taniguchi, H. Du and J. S. Lindsey, PhotochemCAD spectral database (no licence stated)
    11. Max Planck Institute for Brain Research, Spectra viewer (no licence stated)
    12. G. McNamara, PubSpectra (2006; no licence recorded)
    13. M. Drobizhev, N. S. Makarov, S. E. Tillo, T. E. Hughes and A. Rebane, “Two-photon absorption properties of fluorescent proteins,” Nat. Methods 8, 393–399 (2011)
    14. J.-C. Diels and W. Rudolph, Ultrashort Laser Pulse Phenomena, 2nd ed., Academic Press (2006), chapter 1: time–bandwidth product of Gaussian pulses
    15. C. Xu and W. W. Webb, “Measurement of two-photon excitation cross sections of molecular fluorophores with data from 690 to 1050 nm,” J. Opt. Soc. Am. B 13, 481–491 (1996)
    16. W. R. Zipfel, R. M. Williams and W. W. Webb, “Nonlinear magic: multiphoton microscopy in the biosciences,” Nat. Biotechnol. 21, 1369–1377 (2003)
    17. R. S. Molina, Y. Qian, J. Wu, Y. Shen, R. E. Campbell, M. Drobizhev and T. E. Hughes, “Understanding the fluorescence change in red genetically encoded calcium ion indicators,” Biophys. J. 116, 1873–1886 (2019)
    18. S. de Reguardati, J. Pahapill, A. Mikhailov, Y. Stepanenko and A. Rebane, “High-accuracy reference standards for two-photon absorption in the 680–1050 nm wavelength range,” Opt. Express 24, 9053 (2016): Eqs. (1)–(2)
    19. N. S. Makarov, M. Drobizhev and A. Rebane, “Two-photon absorption standards in the 550–1600 nm excitation wavelength range,” Opt. Express 16, 4029 (2008)